TECNIS EyhanceTM Toric II IOL with TECNIS SimplicityTM Delivery System DIU100, DIU150, DIU225, DIU300, DIU375, DIU450 and DIU525
1-Piece toric aspheric lens with ProTec edge design1
• Toric aspheric anterior surface1
• Available in 7 cylinder powers1
• The aspheric anterior surface of the lens corrects spherical aberration to essentially zero1
Engineered for rotational stability
New squared and frosted haptic design provides resistance to rotation2
• Insertion with: TECNIS SimplicityTM Delivery System
• Range available: From 5.0 D to 34.0 D in 0.5 D increments
• Lens diameter/overall length: 6.0/13.0 mm
• Edge design: ProTec frosted continuous 360° posterior square edge
• Material: Soft foldable hydrophobic acrylic with a covalently bound UV absorber
• A-Constant optical/ultrasound3: 119.3*/118.8Ɨ
• Theoretical anterior chamber depth (ACD)(optical/ultrasound): 5.7/5.4 mm
• Refractive index: 1.47 at 35°C
• Haptics: Hydrophobic acrylic
• Haptic design: Haptics offset from optic, Tri-Fix design
• Cylinder/Addition: Range of cylinder powers: +1.0 D, +1.5 D, +2.25 D, +3.0 D, +3.75 D, +4.5 D, +5.25 D
*Value theoretically derived for a typical 20.00 D lens. Johnson & Johnson Vision recommends that surgeons personalize their A-constant based on their surgical techniques and equipment, experience with the lens model and postoperative results.
†Derived from clinical evaluation results of the TECNISTM 1-Piece Platform.
1. TECNIS EyhanceTM Toric II IOL with TECNIS SimplicityTM Delivery System, Model DIU – DfU – INT- Z311515P, Rev. A, Sep. 2020. REF2020CT4293.
2. DOF2019OTH4015 – Study NXTG-103-MER3 – Proof of Concept Study for Next-Generation IOL Models MER003 and MER004. Aug. 28, 2019.
3. Holladay JT. International Intraocular Lens & Implant Registry 2003. J Cataract Refract Surg. 2003; 29:176-197. REF2016CT0151.